Brand: BIOBASE | Category: Scientific & Laboratory Equipment
BIOBASE EN Class II A2 Biological Safety Cabinet BSC-3FA2 BSC-4FA2 is listed in VVON's biosafety cabinets range for Bangladesh. Published models: BSC-3FA2, BSC-4FA2. Compare cabinet class, usable width, exhaust arrangement and installation requirements. For a quotation, send the sample workflow, required cabinet class, working width and room ventilation. Confirm the exact model, included accessories, consumables, delivery and service scope in the offer. Manufacturer specifications below describe the published configuration; options and local availability require confirmation.
Compare each model column separately. Confirm options and included accessories in your quotation.
| Model | BSC-3FA2 | BSC-4FA2 |
|---|---|---|
| External Size(W*D*H) | 1087*775*2265mm | 1383*775*2295mm |
| Internal Size(W*D*H) | 910*600*660mm | 1210*600*660mm |
| Work Surface Height | 750mm | 750mm |
| Max Opening | 520mm(21") | 550mm(22") |
| Tested Opening | Safety height 200mm(8") | Safety height 200mm(8") |
| Airflow Volume / Inflow | 348m³/h(205cfm) | 462m³/h(272cfm) |
| Airflow Volume / Down Flow | 65%: 649m³/h(382cfm) | 65%: 863m³/h(507cfm) |
| Airflow Volume / Exhaust | 35%: 348m³/h(205cfm) | 35%: 462m³/h(272cfm) |
| Front Window | Manual. Laminated toughened glass. Anti UV | Manual. Laminated toughened glass. Anti UV |
| HEPA Filter | Two, 99.995% efficiency at 0.3μm, filter life indicator | Two, 99.995% efficiency at 0.3μm, filter life indicator |
| Filter Guard Type | Aluminum alloy frame | Aluminum alloy frame |
| Noise | EN 12469≤65dB | EN 12469≤65dB |
| Illumination | ≥1000Lux | ≥1000Lux |
| Display | LCD display | LCD display |
| Waterproof Socket | 2, total consumption: ≤500W | 2, total consumption: ≤500W |
| Tap | Water tap*1, Gas tap*1 | Water tap*1, Gas tap*1 |
| Ground Resistance | ≤ 0.10Ω | ≤ 0.10Ω |
| Power Supply | AC230V, 50Hz | AC230V, 50Hz |
| Consumption | ≤550W | ≤650W |
| Alarm | Abnormal airflow velocity, Filter replacement, Front window at unsafe height, High filter pressure alarm | Abnormal airflow velocity, Filter replacement, Front window at unsafe height, High filter pressure alarm |
| Motor | AC motor, 230V acceptable, speed adjustable, high efficiency and low power consumption | AC motor, 230V acceptable, speed adjustable, high efficiency and low power consumption |
| Material | Working area: 304 stainless steel Frame and decorative plate: cold-rolled steel with anti-bacteria power coating | Working area: 304 stainless steel Frame and decorative plate: cold-rolled steel with anti-bacteria power coating |
| UV Lamp | 20W | 30W |
| UV Lamp | Germicidal UV lamp, Emission of 253.7 nanometers for most efficient decontamination, UV timer, UV life indicator | Germicidal UV lamp, Emission of 253.7 nanometers for most efficient decontamination, UV timer, UV life indicator |
| LED Lamp | 10W*2 | 18W*2 |
| Standard Accessory | LED lamp*2, UV lamp*2, Base stand, SS water tap*1, SS gas tap*1, Drain valve, Waterproof sockets*2, Stainless steel I.V. bar with 10 hooks, Remote control | LED lamp*2, UV lamp*2, Base stand, SS water tap*1, SS gas tap*1, Drain valve, Waterproof sockets*2, Stainless steel I.V. bar with 10 hooks, Remote control |
| Optional Accessory | Armrest, Airflow tester, Formalin fumigation sterilizer, Infrared sterilizer,; Electric height adjustable base stand, Motorized front window | Armrest, Airflow tester, Formalin fumigation sterilizer, Infrared sterilizer,; Electric height adjustable base stand, Motorized front window |
| Package Size(W*D*H) | 1240*1060*1900mm | 1540*1080*1900mm |
| Gross Weight | 270kg | 340kg |
Manufacturer catalogue checked 2026-09-28.
Request an itemised quotation covering the proposed configuration, availability, lead time and agreed supply and support scope.
Share what you know. You can start with your name and a phone number or email; final specifications can follow.
Not sure which configuration you need? Help me choose
Contact Vvon Technologies Limited about BIOBASE equipment requirements in Bangladesh.
Phone: +8801805987890 | Email: info@vvon.com.bd
Back to Scientific & Laboratory Equipment
Biosafety Cabinets by BIOBASE. Compare cabinet class, usable width, exhaust arrangement and installation requirements.
Biosafety Cabinets by BIOBASE. Compare cabinet class, usable width, exhaust arrangement and installation requirements.
Volume Exhaust: 35%: 230m³/h(135cfm). Compare cabinet class, usable width, exhaust arrangement and installation requirements.