Brand: Delta Elektronika | Category: Scientific & Laboratory Equipment
An Ethernet interface between a TCP/IP network and a Delta power supply, ordered as a factory fitted board or as an external module. It provides a web interface, SCPI control, status read back, eight isolated digital inputs and outputs and a stored waveform sequencer.
| Parameter | Value |
|---|---|
| Programming and monitoring resolution | 16 bit |
| Linearity error | plus or minus 2 LSB programming, plus or minus 1 LSB monitoring |
| Temperature coefficient | 10 ppm per °C |
| Sequencer | 25 freely programmable sequences of 2000 steps each |
| Input voltage (external module) | 98 to 264 V AC (48 to 62 Hz) |
| External module | 89 mm wide, 86 mm high, 119 mm deep, 0.7 kg |
Contact Vvon Technologies Limited about Delta Elektronika equipment requirements in Bangladesh.
Phone: +880 1711-738207 | Email: info@vvon.com.bd
Back to Scientific & Laboratory Equipment
An RS232 interface for analogue programmable Delta supplies, available as a factory fitted board or an external module. It sets output voltage and current and reads back voltage, current and status at 14 bit resolution, with up to 15 units on one bus.
An external carrier for Anybus CompactCom M40 inserts that gives analogue programmable power supplies a fieldbus connection. It translates between the analogue interface and the chosen protocol, and can also work as an analogue to USB interface using SCPI commands.
An isolation amplifier that adds galvanic isolation to the analogue programming connector, sold as a factory fitted card or an external module. Signal levels are selectable and it isolates voltage, current and status lines to avoid earth loops and common mode faults.