CBCT-150

Brand: Leeds Test Objects | Category: Nuclear Medicine & Imaging

A PMMA cuboid phantom for constancy testing of cone beam CT and 3D fluoroscopy systems to DIN 6868-150. Rows of small air holes of decreasing diameter give a direct measure of spatial resolution, and a tripod allows universal positioning.

Technical Specifications

ParameterValue
SpecificationPMMA cuboid 120 x 120 x 60 mm
SpecificationRows of 4 air holes of 1.3, 1.0, 0.90, 0.90, 0.90, 0.60 and 0.50 mm diameter
SpecificationSpatial frequencies 0.38, 0.50, 0.56, 0.63, 0.71, 0.83 and 1.0 mm respectively
SpecificationTripod for universal positioning
SpecificationConforms to DIN 6868-150