Skip to main content

Collimator selection for gamma camera imaging: matching the hardware to the isotope and the study

Low energy high resolution, general purpose, high sensitivity, pinhole and fan beam collimators do different jobs. How to choose the right one for each study, and how to store and handle a component that cannot be repaired locally.

A department in a divisional hospital takes delivery of a dual head gamma camera with one pair of low energy high resolution collimators. Bone, renal and thyroid work runs well for a year. Then an endocrinologist asks for post-therapy I-131 imaging, and the department discovers that the only collimators in the building will let 364 keV photons pass straight through the septa. The images come back with a haze that no reconstruction setting will remove, and the fix is an import order with a lead time measured in months.

Collimators are the least glamorous item on a gamma camera purchase order and the one that most often gets trimmed at the negotiation stage. They also determine the resolution and count rate of every study the department will ever do. Nothing downstream, no reconstruction algorithm and no detector technology, recovers information that the collimator threw away.

What a collimator does, in one paragraph

A gamma camera detector records where a photon landed but not the direction it came from. Without a collimator, every point in the patient illuminates the whole detector and the image is a uniform blur. A parallel hole collimator is a lead honeycomb placed in front of the detector: photons travelling roughly parallel to the holes pass through, and photons arriving at an angle are absorbed by the septa between them. The image is formed by rejection. Only a very small fraction of emitted photons make it to the crystal, which is why nuclear medicine acquisitions are measured in minutes rather than milliseconds.

The trade-off that governs every choice

Two properties compete. Spatial resolution improves with narrower holes and longer septa, because the accepted cone of directions is narrower. Sensitivity, the fraction of emitted photons recorded, improves with wider holes and shorter septa, for the same reason in reverse. The relationship is not linear: sensitivity varies roughly with the square of resolution, so a collimator that halves the accepted angle costs you far more than half your counts. That is the whole design space, and every named collimator type is a chosen point within it.

Two practical consequences follow. First, resolution degrades as the source moves away from the collimator face, so patient contact matters. A technologist who leaves a five centimetre gap because the couch is easier to position that way has thrown away more resolution than any purchasing decision could recover. Second, adding count time can partly compensate for a low sensitivity collimator, but nothing compensates for a low resolution one.

The low energy family: three points on the same curve

Almost all routine work uses Tc-99m at 140 keV, so most departments live inside the low energy family. The three common members are the same idea at different settings.

CollimatorBiasTypical use
Low energy high resolution (LEHR)Resolution over countsBone scans, renal cortical imaging, most planar and SPECT work with Tc-99m where the patient can lie still long enough
Low energy general purpose (LEGP or LEAP)BalancedMixed lists, dynamic studies where frames are short, larger or restless patients, departments that own only one pair
Low energy high sensitivity (LEHS)Counts over resolutionFirst pass and flow studies, very short frames, paediatric work where movement is inevitable and speed matters more than fine detail

If a department can afford one pair only, LEHR is usually the right single choice for a general list, with the understanding that fast dynamic work will be noisier than it needs to be. If the department expects a lot of flow and first pass studies, the general purpose collimator is the safer single purchase. Buying only high sensitivity collimators to make lists faster is a false economy that shows up as unreportable spine images.

Medium and high energy: the isotope decides, not the study

Septal thickness has to scale with photon energy, because higher energy photons penetrate lead more readily. Push an isotope beyond what a collimator was designed for and photons travel through the septa rather than the holes. The result is a characteristic star or haze pattern, a loss of contrast, and counts appearing in places where there is no activity.

Special geometries: pinhole, fan beam and the rest

Parallel hole collimators map the patient one to one onto the detector. Where the organ is much smaller than the detector, that wastes most of the field of view, and a different geometry does better.

Pinhole

A pinhole collimator is a lead cone with a small aperture at the tip. It magnifies a small structure onto the full detector, which is why it remains the best planar tool for thyroid imaging, parathyroid localisation, paediatric hips and small joints. Most pinhole collimators accept interchangeable aperture inserts of different diameters, trading resolution against counts in the same way as a parallel hole collimator. Two things trip up new users: sensitivity falls off steeply as the organ moves away from the aperture, so distance has to be reproducible if you want comparable images; and the image is geometrically distorted towards the edges, so a nodule position near the edge of the field is not where it looks.

Fan beam and converging geometries

A fan beam collimator has holes that converge in one direction and stay parallel in the other, magnifying the brain onto a detector designed for a torso. Brain SPECT with a fan beam gains counts and resolution together, which is unusual and welcome. The trade is truncation: anything wider than the converged field is cut off, so it is a brain tool and not a general one. Converging and diverging cone geometries and slant hole designs exist for particular applications and are rarely worth specifying unless a named clinical service requires them.

A working table for a general department

StudyIsotopeCollimator
Whole body bone scan and bone SPECTTc-99m MDP or HDPLEHR
Renography and other dynamic renal workTc-99m DTPA, MAG3 or ECLEHR or LEGP, depending on frame length
Thyroid uptake and nodule imagingTc-99m pertechnetate or I-123Pinhole for detail, LEHR for a survey view
Post-therapy and whole body iodine imagingI-131High energy
Infection and inflammation imagingGa-67 or labelled leucocytes with In-111Medium energy
Myocardial perfusion SPECTTc-99m sestamibi or tetrofosminLEHR
Gated blood pool and first passTc-99m labelled red cells or pertechnetateLEGP for gated work, LEHS for first pass
Brain perfusion SPECTTc-99m HMPAO or ECDFan beam where available, otherwise LEHR

Handling and storage: the damage nobody admits to

A collimator is a slab of lead with a fragile face. The honeycomb is thin lead, the cover over it is a light foil or composite skin, and both crush easily. A dent is not cosmetic. It bends septa, and that produces a fixed non-uniformity which appears on every subsequent image and on every flood. There is no local repair. A damaged collimator is an import order and a hole in your service list.

What to write into the tender

Collimator specifications are where a low bid quietly becomes an expensive machine. A line reading collimators as required is an invitation to supply the minimum.

  1. Name each collimator type and the quantity per detector head, in a table, in the technical specification.
  2. State the isotopes and clinical studies the department intends to run in the first three years, including any therapy service, so that medium and high energy needs are visible at evaluation.
  3. Require the collimator exchange trolley or cart, and state whether exchange is manual or automatic. On a dual head system this is a daily operation, not an occasional one.
  4. Require one storage cradle or stand per collimator, and confirm the room has floor space for them. Collimator stands are often left off drawings and then have nowhere to live.
  5. Require an extrinsic flood source holder compatible with each collimator, and confirm how flood sources will be obtained and replaced in Bangladesh.
  6. Ask for the lead time and landed cost of a replacement collimator in writing at tender stage, while you still have negotiating position. After installation, that quotation is whatever the supplier says it is.

Back to all Insights